17.Refer to
Figure
19. Locate W34 semi-rigid coax cable (E4410-20164) in the kit. Connect the
cable between the center port of Switch 1 (the lower switch) and J9 of the A13 Front End
Assembly. Torque to 10 inch-pounds.
18.Locate W31 semi-rigid coax cable (E4410-20161) in the kit. Connect the cable between the
center port of Switch 2 (the upper switch) and J3 of the A11 Low Band Switch. Torque to
10 inch-pounds.
Figure 19
Install W34 and W31
Installation Note N9020-90253
Installation Procedure
31