14.4 Auto Print

Using Auto Print allows measurement results to be automatically output via the RS-
232C interface without pressing the [PRINT] key individually with every measure-
ment. An issue will be made after standstill control ( ) is complete. The next meas-
urement is done after removing the items from the weighing plate and the display is
returning to a value within the ±3-fold of the zero range.
Important:
 This mode can be applied to any unit.
 The initial load of the balance must be kept within the ± 5-fold of the ze-
ro range.
 Weight of sample ≥ 10 x zero range (zero range = 0.25 e)
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
Repeatedly press the [CAL] key until the current "AtPrt:**" set-
ting appears.
(Example)
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
" AtPrt-on"
" AtPrt-oF"
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
To exit the function, press the [ON/OFF] key.
Brief actuation of ON/OFF key:
Back to previous menu.
Long actuation of ON/OFF key:
Back to weighing mode. Whilst the auto print function is active
the display keeps showing the [AP] symbol.
ABT-BA-e-1312
Function activated
Function deactivated