15.3 Remote control instructions

Attention:
Inputting characters and command codes not shown here into the balance may not
only alter the previous settings but may also impair proper measurement.
If by mistake characters or commands not shown here are entered into the balance,
immediately unplug the power supply cable and wait about ten seconds before plug-
ging again.
Command Code
D01
Continuous output
D05
D06
1 time output with stability
D07
D08
1 time output at stability
D09
Q
ON/OFF switching
T
TS
Taring after stability wait
C18
Measurement span calibration
+
Measurements in add-up mode
R
mg
PERCENT
Percent determination
PCS
Parts counting
CT
SDENCE
Density of solids
LDENCE
Density of liquids
%
G
g, % switching
- g
g unit removal
- mg
mg unit removal
Deleting percentage determina-
- PERCENT
- PCS
Piece counting removal
- CT
ct unit removal
- SDENCE
Solid specific gravity removal
- LDENCE
Liquid specific gravity removal
ABT-BA-e-1312
Function
The balance continuously outputs every 110ms .
1 time output
Corresponds to [PRINT] key
Auto Print
See 14.4
The status of the stability mark is appended to the head
of the data with output.
information
S: when the standstill display appears
U: when standstill display does not appear
After command input, the data are output at the first
appearance of the standstill display.
Halt output
Auto Print and continuous output halted
Switches between standby status and measurement
status.
Taring
Corresponds to the [O/T] key
After command input, taring is done at the first appear-
ance of the stability mark.
See ch. 14.5
Total reset
All application measurements terminated and reset
mg unit
ct unit
100% setting
tion t