Recipe Mode - KERN ABT Series Operating Instructions Manual

Analytical balance
Hide thumbs Also See for ABT Series:
Table of Contents

Advertisement

Available languages

Available languages

Printout:
CMP001=
CMP002=
CMP003=
TOTAL=
Note:
Only one function at a time may be activated (on) when working with func-
tions auto print, add-on and recipe.
Whilst the function is active, automatic adjustment (PSC/Clock-CAL) does
not take place. A flashing
required.

14.6 Recipe mode

The formula function allows to add on various components of a mixture. The total
weight of all components can be invoked to check (separate memory for the weight
of tare vessel and recipe components).
Activate function "FormU: on" (see chap. 14.5 Add up):
Whilst the recipe function is active, the display keeps showing the "FormU -on" sym-
bol [
M].
Operation
When using a container, place this on the weighing plate
and tare.
Weigh first component.
When standstill control ( ) is complete, the weighing
value of the first component (CMP001) is issued to the
optional printer by pressing the [PRINT] key. The dis-
played value is added into the total adding memory. Af-
terwards automatic taring will take place.
Weigh second component
When standstill control ( ) is complete, the weighing
value of the second component (CMP002) is issued to
the optional printer by pressing the [PRINT] key. The
displayed value is added into the total adding memory.
Afterwards automatic taring will take place.
62
---ADDON Mode---
5,0003 g
5,0017 g
5,0010 g
15,0030 g
symbol, however, indicates that adjustment is
Display
0.0000 g�
0.5361 g
0.5361 g
0.0000 g
0.5422 g
0.54221 g
0.0000 g
ABT-BA-e-1312