Download Print this page

Changing The Scale Place Of Operation; Cleaning And Maintenance - KERN IEX Safety-Usage Manual

Hide thumbs Also See for IEX:

Advertisement

zero indication
stable result
weight unit
kg
6

Changing the scale place of operation

While moving a scale in another place of operation, all safety
measures listed in point 5 of this instruction should be taken.
7

Cleaning and maintenance

Unplug the power adapter IEX-A01 from mains.
Check wether the indicator´s electrostatic grounding is correctly plugged and is
not defected.
Clean the scale and the housing with a slightly moistened cloth. You can use
typical domestic cleaning agents.
For clearing an overlay of the indicator do not use agents that can
mechanically scratch the surface. It includes an internal conductive layer that is
important for save operation.
Platform levelling feet and shock absorbers are made of rubber. Within the
hazardous area they need to be cleaned with a slightly moistened cloth. Moisture
protects against gathering electrostatic charges.
While clearing the platform do not damage the masking springs. They are for
gathering electrostatic charges from the weight pan. They should be ~5mm above
the shock absorbers.
Cleaning should be performed when there are no explosive gases.
There is forbidden to us concentrated acids, bases, solvents or alcohol for
cleaning the device.
There is also forbidden to use compressed air and high pressure washing
equipment.
10
IEX-BASV-e-1711

Advertisement

loading