Agilent Technologies OmniBER OTN Series Performance Verification Manual page 60

Communications performance analyzer
Hide thumbs Also See for OmniBER OTN Series:
Table of Contents

Advertisement

Verification.book Page 57 Monday, January 19, 2004 1:40 PM
OmniBER OTN
Verification Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OC-192/STM-64 Rate (J7230B and J7232A Opt 108 Only)
1 Press <Menu>, choose System > Stored Settings then press
<Select>. Select Default then Recall.
2 Connect the test setup as shown below.
Optical In
Optical Out
8156A Optical Attenuator
9.95 Gb/s Optical Output
3 Set the optical attenuator to:
ATTEN 15 dB
WAVELENGTH 1550 nm
CAL=0
ENB ON
Set the optical power meter to
WAVELENGTH 1550 nm
Connect the optical attenuator between the power meter and
the OmniBER OTN's Optical Out port (ensure that all
connections are tight and that the cable has no twists).
8163/81633A Optical Power Meter
step 6