Verification.book Page 42 Monday, January 19, 2004 1:40 PM
1
Performance Tests
42
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
11 Check the Extinction Ratio is greater than 8.2 dB and record
the measured value.
OC-192/STM-64 Eye Mask
10.71 Gb/s (OTU-2) Power Level and Eye Mask Tests (Option 112
only)
12 Set the Transmitter display as shown below.
Press <Menu>, choose Tx/Rx > Transmitter Settings >
Physical then press <Select>. Set the Signal Mode to OTN
and the Signal Rate to OTU-2.