2M Balanced Negative Pulse
2 Mb/s Unbalanced Output Mask Tests
1 Press <Menu>, select Tx/Rx > Transmitter and set the
Signal Mode to PDH/DSn. Now set the Signal Rate to
2Mb/s, Termination 75 ohms (Unbal).
2 Connect the 2 Mb/s BAL OUT port to the Oscilloscope CHAN 1,
using the E2622A Termination Adapter Probe.
1
Performance Tests
69