DS1 Output Mask Test
1 Recall the OmniBER OTN's default settings as shown on
"Recall Default Settings"
2 Press <Menu>, select Tx/Rx > Transmitter Settings and set
the Signal Mode to PDH/DSn. Now set the Signal Rate to
DS1 as shown in the example below.
3 Connect the OmniBER OTN's DS1/2 Mb/s OUT port to the
Oscilloscope using the E2621A Termination Adapter Probe,
set probe switch to 100 Ω.
54810A Oscilloscope
DS1/2 Mb/s Balanced O/P
Bantam to Bantam Cable
Performance Tests
on page 11.
CH1
Set
E2621A
to
Bal/Unbal
100Ω
Adapter
1
65