Service Settings And Checks; The Service Menu; Activating The Service Menu - Physiomed PHYSIOTHERM-S Service Manual

Hide thumbs Also See for PHYSIOTHERM-S:
Table of Contents

Advertisement

4 Service Settings and Checks

4.1 The Service Menu

4.1.1

Activating the Service Menu

• Call up the basic settings menu (keep the data selector depressed while switching
on the unit).
• Turn the data selector until the service code line is highlighted.
• Press the data selector. Set the service code to 1310.
• Select Esc, press the data selector and press Esc again. The service menu is activated
(see image on the next page).
• Put a bucket filled with water between the plate electrodes.
• Set HF – power output to 150 Watt / cont.
• Fill in the measured values:
U1: _______
U2: _______
U3: _______
The displayed values are important assessment criteria for our service depart-
!
ment
(You can print or copy this page to fill in the values.)
EFFECTIVE POWER
0.0W
MATCHING
U1= 000
U2= 000
U3= 000
LT
RT
Display
Significance
U1
Forward power
U2
Reverse power
U3
Adaptation value
T1
Temperature of preliminary amplifier
T2
Temperature of power amplifier
T3
Temperature of operating panel
HF1
Output power signal generator
HF2
Output power preliminary amplifier
HF3
Output power amplifier
PHYSIOTHERM-S- Service Manual
T1: _______
HF1: _______
T2: _______
HF2: _______
T3: _______
HF3: _______
Output Power
Time
Mode
T1= 075
T2= 084
T3= 105
Hz
Service Settings and Checks
0 W
60 min
HF1= 000
HF2= 000
HF3= 000
MODE
min.
Display
Significance
LT
Left hand rotation (tuning capacitor)
RT
Right hand rotation (tuning capacitor)
Hz
Frequency (pulsed mode)
Mode
Operating mode
min
Treatment duration
W
Output power
W
21