Table 6-1. Battery Power Status - Covidien Nellcor GR101704 Service Manual

Bedside respiratory patient monitoring system
Table of Contents

Advertisement

Battery Fuel Gauge
Caution:
Should a low battery alarm sound, connect the monitoring system to an AC
power source and then silence the alarm by pressing ALARM SILENCE. If the
monitoring system is operated on an AC power source with a depleted
battery and AC power is subsequently lost, the monitoring system will shut
down immediately.
The monitoring system runs on an internal battery when not connected to an
AC power source. A battery fuel gauge displays the remaining battery power.
When connected to AC power, the battery fuel gauge displays a lightning bolt
while charging and at full charge.
Reference
p. 5-6.
Note:
The battery is recyclable. Do not dispose of the battery by placing it in the regular
trash. Dispose of the battery in accordance with local guidelines and regulations or
contact Covidien to arrange for disposal.
Note:
As the battery is used and recharged over time, the amount of time between the onset
of low battery alarms and the monitoring system shut-off may become shorter.
Message
None
BATTERY LOW
BATTERY CRITICALLY
LOW
1. The levels listed are based on a new battery. Continued battery charge and discharge eventually reduces capacity. For
example, a battery two years old may provide only 75% of the capacity of a new battery.
Service Manual
Connection to an AC Power
Table 6-1.  Battery Power Status
Color
Green
Normal Status — Indicates 15-100% (approximately 15
Yellow
Low Status — Indicates 5-14% (approximately 15 minutes)
Red
Critical Status — Indicates 1-4% (approximately 5 minutes)
Source, p. 5-3. Reference
1
Power Charge Status
minutes to 6 hours) battery capacity remains.
battery capacity remains.
battery capacity remains.
Battery Power
Usage,
Power
6-3