Figure 12-7. Parameter Module Tab Release; Figure 12-8. Parameter Module Assembly Removal - Covidien Nellcor GR101704 Service Manual

Bedside respiratory patient monitoring system
Table of Contents

Advertisement

Repair
Gently press inward against the parameter module assembly through the two
4.
openings, releasing it from the backplane. Expect minor resistance until it releases.
Remove and set aside the parameter module assembly.
5.
Gently rotate the monitoring systemto the upright position.
6.
12-12
Figure 12-7.  Parameter Module Tab Release
Figure 12-8.  Parameter Module Assembly Removal
Service Manual