INSTALLATION
2

2.8.7 How to turn or remove the signal converter

The converter turns 360°, but we recommend that the tag hole on the housing points to the
nearest tank wall. For more data, refer to
converter can be removed from the process connection assembly under process conditions.
Figure 2-34: How to turn or remove the signal converter
1 Tool: 5 mm Allen wrench (not supplied) for the lock screw on the signal converter
2 Cover for the wave guide hole on top of the process connection assembly (not supplied)
CAUTION!
Do not loosen the 4 socket head screws on the process connection assembly.
If you remove the housing, put a cover on the wave guide hole on top of the process connection
assembly. Make sure that the wave guide hole is clean and dry.
When the housing is attached to the process connection assembly, tighten the lock screw with
the 5 mm Allen wrench 1.
27
Recommended mounting position
www.drexelbrook.com
DR5200
on page 15. The
DR5200QS-LM, EDO#6-13-112 - Issue #3