DR5200
CAUTION!
If there are parasitic signals, the device will not measure correctly. Parasitic signals are caused
by:
Objects in the tank.
•
•
Sharp corners that are perpendicular to the path of the radar beam.
Sudden changes in tank diameter in the path of the radar beam.
•
Do an Empty Spectrum recording (refer to Operation
Requirements for flange connections
Figure 2-28: Flange connection
Equipment needed:
• Device
• Flange gasket (not supplied)
• Wrench (not supplied)
• Make sure the flange on the nozzle is level.
• Make sure that you use the applicable gasket for the flange dimensions and the process.
• Align the gasket correctly on the flange facing of the nozzle.
• Lower the antenna carefully into the tank.
• Make sure that you point the device in the correct direction. Refer to "Point the device in the
correct direction" in this section.
• Tighten the flange bolts.
i
Refer to local rules and regulations for the correct torque to apply to the bolts.
DR5200QS-LM, EDO#6-13-112 - Issue #3
Operation) to remove parasitic signals with a filter.
Operation
Operation
www.drexelbrook.com
INSTALLATION