Ametek DR5200 Quick Start Manual page 13

Radar (fmcw) level transmitter for liquids in storage and process applications
Hide thumbs Also See for DR5200:
Table of Contents

Advertisement

INSTALLATION
2
Metallic Horn and Wave Guide: compact and remote versions
Ambient temperature / flange temperature, flange and threaded connection, in °C
Figure 2-15: Ambient temperature / flange temperature, flange and threaded connection, in °C
Ambient temperature / flange temperature, flange and threaded connection, in °F
Figure 2-16: Ambient temperature / flange temperature, flange and threaded connection, in °F
1 Maximum ambient temperature, °C
2 Maximum flange temperature, °C
3 Maximum ambient temperature, °F
4 Maximum flange temperature, °F
5 Compact and remote versions of the device with the Metallic Horn or Wave Guide antenna
6 Compact version of the device with the Metallic Horn or Wave Guide antenna
INFORMATION!
When the process temperature is -50
de-rating for the ambient temperature. The minimum ambient temperature is -35
13
C / -58
F and the device has an EPDM gasket, there is a
°
°
www.drexelbrook.com
DR5200
°
C / -31
°
DR5200QS-LM, EDO#6-13-112 - Issue #3
F.

Advertisement

Table of Contents
loading

Table of Contents