Ametek DR5200 Quick Start Manual page 46

Radar (fmcw) level transmitter for liquids in storage and process applications
Hide thumbs Also See for DR5200:
Table of Contents

Advertisement

DR5200
Screen
DR5200QS-LM, EDO#6-13-112 - Issue #3
Steps
Push [> > > > ] to make a selection from the
available spectra.
[^ ^ ^ ^ ] to confirm.
[> > > > ] to set to "Yes". [ ] to set to "No".
www.drexelbrook.com
OPERATION
Description
Recording in progress
Recording in progress
Recording in progress
Recording in progress
Reading in progress
Reading in progress
Reading in progress
Reading in progress
The device does a scan for objects that do
not change their vertical positions in the
tank (heating tubes, agitators, fuel
assemblies etc.) and records the data.
This shows a set of spectra (signals found in
the tank). It also gives the amplitude and
location of each signal.
Do you want to save the spectrum?
Do you want to save the spectrum?
Do you want to save the spectrum?
Do you want to save the spectrum?
Set to "Yes" to use the data. Set to "No" to
erase the data.
4
46

Advertisement

Table of Contents
loading

Table of Contents