With holes or recesses
z
When mounted on a wall
z
Attach the sensor head so that the window (emitting/receiving surfaces) is parallel to the detection target.
Adjust the target so that the spot aligns with the detection position, and check whether the status LED lights
up orange or green.
Caution
■ Warmup
Wait approx. 30 minutes after turning the power ON before using the product. Measurements performed
directly after turning ON the power will gradually drift (change) over time.
■ Effects of Ambient Temperature
Changes in ambient temperature can cause measurement errors. Ensure a constant ambient temperature
at all times. If the ambient temperature changes abruptly, approx. 60 minutes is required for the
temperature inside the device to become uniform.
■ Effects of Dirt and Dust
Measurement errors may occur due to dirt or dust as well as water or oil.
• Should the glass cover or the surface of the measurement target become dirty, clean it by blowing it with
clean air.
• If the glass cover becomes excessively dirty, wipe with a soft cloth.
Mounting the Head
1
1-9