7.
Set B value and press [Enter] key.
TRANSITION
Level B=2.0A
8.
Start dynamic test and operate
[Enter] key.
TRANSITION
On
9.
Enter the dynamic test mode.
10.00V 0.0A
0W
10. Press [On/Off] key to open input and press [Shift] + . (Trigger) key.
The load will switch after receipt of every trigger signal. The load will switch
between A and B value for one time. Time of operations is shown at bottom
right.
11. Press [CC]/ [CV]/ [CR]/ [CW] key or any composite function key to exit
dynamic test function. Repeat Steps 1-10 to continue parameter setting
and operation of dynamic test.

3.11 OCP test function

The IT8300 series electronic load is provided with overcurrent protection test
function (OCP). Under OCP test mode, when input voltage reached Von value,
delay for a while for the electronic load to latch. Ascend value by step value at
regular interval. At the same time, check the load input voltage and judge
whether it is higher than standard voltage value. If higher, it indicates that OCP
does not occur. Repeat current stepping operation till the load operates to the
cutoff current. At the same time, check whether the cutoff current value is within
target scope. If yes, PASS the test, otherwise FAULT. If lower, it indicates that
OCP has occurred. Check whether the existing current value is within target
scope again. If yes, PASS the test, otherwise FAULT.
Operating steps:
1.
Press [Shift] + [CC] (OCP) keys to enter OCP test function setting screen.
Run
Recall
OCP
TEST
Edit
Off
0 TRAN
OCP TEST
OCP TEST
Recall OCP File=1
OCP TEST
1:Voltage on level=0.00V
2:Voltage on Delay=0.00S
3:Current Range=0.0A
4:Start Current=0.0A
5:Step Current=0.0A
6:Step Delay=0.00S
7:End Current=0.0A
8:OCP Voltage=0.00V
9:Max Trip Current=0.0A
10:Min Trip Current=0.0A
Copyright © ITECH Electronics Co., Ltd.
key. Move to on and press
Run OCP test documents.
Recall OCP test documents (1-5).
Set Von voltage value.
Set Von voltage delay time.
Set working current range.
Set initial current value.
Set step current value.
Set step delay time.
Set cutoff current value.
Set OCP value.
Set overcurrent range (maximum
value).
Set overcurrent range (minimum
Function and Features