1.4 Connecting Test Cables (Optional)

Test cables are not standard accessories of the instrument. Please select
optional red and black test cables for individual sales based on the maximum
current value. For specifications of test cables and maximum current values,
refer to "Specifications of Red and Black Test cables" in "Appendix".
Before connecting test cables, be sure to switch off the instrument.
Power switch is in Off position. Otherwise, contact with input
terminals in rear panel may cause electrical shock.
To avoid electrical shock, before testing, please make sure the rating
values of the testing cables, and do not measure the current that
higher than the rating value.
Always use test cables provided by ITECH to connect the equipment.
If test cables from other factories are used, please check that the test
cable can withstand maximum current.
Test cable connection is given below taking local measurement as example.
For details of local and remote measurements, refer to "Functions of Rear
Panel Terminal".
1.
Before connecting the test cables, be sure that the instrument Power is in
Off position.
2.
Unscrew the screws of the input terminals and connect the red and black
test cables to the input terminals. Re-tighten the screws.
When maximum current that one test cable can withstand fails to meet the
current rated current, use several pieces of red and black test cables. For
example, the maximum current is 1,200A, then 4 pieces of 360A red and
black cables are required.
3.
Directly connect the other end of the red and black cables to the DUT
terminal.
Copyright © ITECH Electronics Co., Ltd.
Inspection and Installation
7