3.14 Configuration save function

The electronic load can save some commonly-used parameters in 100-group
NVM for convenient and fast usage. The parameters to be saved includes
working mode, voltage, current, etc. The [Shift] + 4 (Save) keys can be used
for saving parameters. The [Shift] + [Enter] (Recall) key is for quick invoking.
Operation steps
If the operator needs to save configured parameter values for direct recall in
follow-up operation, refer to the steps below:
E.g., power supply 6V and 3A. The electronic load works under constant
current (CC) 1A. Save "CC 1A" in register 9 for recall.
SAVE
1.
Set parameters and press [Shift] + 4 (Save) keys to save data. Press 9 key
(to select in which group the data is to be saved).
5.89V
5W
2.
Press [Enter] key.
5.89V
5W
RECALL
1. Press [Shift] + [Enter] (Recall) key and press 9 (to select from which
Memory function
When you want to recall the data saved in the memory, you should set memory
group in the system menu first.
Group 0 means you can recall data saved in 1-10 groups. Group 1 means you
can recall data saved in 11-20 groups. Group2-Group 9 can be concluded in the
same manner.
capacity. Press [Enter] key. When set battery capacity is reached, the
load input status will automatically be OFF.
Type II: Press [Shift] + 5(Battery) keys and select Voltage. Press
[Enter] key and the VFD displays Stop Voltage=V. Set stop voltage
and press [Enter] key to start discharge test. When the battery voltage
drops to stop voltage, the load input status will automatically be OFF.
Type III: Press [Shift] + 5(Battery) key and select Timer. Press [Enter]
key and the VFD displays Stop Timer=S (Max. 99999S). Set discharge
time. When the set stop time is reached, the load input status will
automatically be OFF.
0.9A
Save 9
0.9A
CC=1.0A
group the data is recalled).
5.89V
0.9A
5W
Recall 9