Battery Discharge Test Function - ITech IT8300 Series User Manual

Regenerative dc electronic load
Hide thumbs Also See for IT8300 Series:
Table of Contents

Advertisement

2.
Press [Shift] + . (Trigger) key to start OPP test. If within range, PASS the
test and the board will display as follows:
9.99V
1W
If not, there is FAULT and the board will display as follows:
9.99V
1W
3.
End test. Press [ESC] to return to setting screen. Press [ESC] again to
exit.

3.13 Battery discharge test function

In the IT8300 series electronic load, constant current mode is applied for
discharge test with programmatic setting of stop voltage/capacity/discharging
time If stop voltage is set as the stop condition, the system determines whether
the battery is about to reach the set threshold value or unsafe status when the
battery voltage is low, and if yes, an automatic stop will be activated. During test,
the operator can observe voltage, discharging time and discharged capacity of
battery. Battery discharge test is a necessary step before battery replacement
as it can reflect reliability and remaining life of battery.
Press [Shift] + 5(Battery) keys to enter function testing screen of battery
discharge test.
STOP
CONDITION
Operation method:
1.
Press [On/Off] key to cut off load input status. Connect the battery to be
tested. Under CC mode, press [Shift] + 5(Battery) keys to enter battery
discharge function menu and perform tests based on stop conditions.
Whichever stop condition is satisfied, the test will stop.
2.
Set discharge stopping conditions:
10:Min Trip Power =0W
Save OPP File=1 (1-5)
0.7A
49W
PASS
0.7A
48W
FAULT
Note
If the set OPP voltage value is higher than the power voltage value, the OPP will fail to
operate and the board will display as follows:
9.99V
0.7A
1W
1W
FAULT
Stop Voltage=0.00V
Stop Capability=0.0Ah
Stop Timer=0S
Type I: Press [Shift] + 5(Battery) keys and select Capability. Press
[Enter] key and the VFD displays Stop Capability =Ah. Set battery stop
Copyright © ITECH Electronics Co., Ltd.
value).
Set overpower range (minimum
value).
Save OPP test documents.
STOP
STOP
STOP
Set stop voltage.
Set battery stop capacity.
Set discharge time.
Function and Features
21