Dimensions Required; Weights, - Beckman Coulter Biomek NXP Preinstallation Manual

Laboratory automation workstation
Hide thumbs Also See for Biomek NXP:
Table of Contents

Advertisement

Preparing the Site
Work Area Minimum Dimensions
Table 1.2 Dimensions Required
P
Biomek NX
Minimum bench necessary to adequately
hold the Biomek NX
Rear access clearance
Side access clearance
a. The dimensions for the lab bench in the above table do not include space for the automation controller or additional accessories. A lab bench
with shelving for accessories underneath the instrument is desirable.
1-4
Component
instrument base unit
P
a
instrument
Minimum Dimensions Required
• NX-MC — 36 in. (L) x 31 in. (W) x 34 in. (H); 91.44 cm (L)
x 78.74 cm (W) x 86.36 cm (H)
• NX-S8 — 36 in. (L) x 31 in. (W) x 44 in. (H); 91.44 cm (L)
x 78.74 cm (W) x 111.76 cm (H)
25 in. (L) x 24 in. (W)
63.5 cm (L) x 61 cm (W)
CAUTION
P
The Biomek NX
instrument will overhang the edges of
a 25" x 24"(63.5 cm x 61 cm) bench. Make sure there
are no obstacles that will interfere with placement of
the instrument.
CAUTION
Make sure the bench can support the total installed
weight of the system. Refer to
the total weight of the system.
4 in. (10.2 cm)
12 in. (30.5 cm)
Table 1.3
to determine
B20063AC