Electrical Requirements; Figure 2-1. Dimensions Of The Optima Tlx Ultracentrifuge - Beckman Coulter Optima TLX Instruction Manual

Preparative ultracentrifuge
Table of Contents

Advertisement

Preinstallation Requirements

ELECTRICAL REQUIREMENTS

2-2
If it is necessary to move the ultracentrifuge, maintain the following
conditions.
• Select a location away from heat-producing laboratory equipment,
with sufficient ventilation to allow for heat dissipation.
• Position the ultracentrifuge on a level surface, such as a sturdy
table or laboratory bench that can support the weight of the
centrifuge (81 kg/178 lb) and resist vibration. (We recommend
placing the instrument at least 5.1 cm (2 in.) from the edge of the
laboratory bench.)
• In addition to space for the ultracentrifuge itself (see Figure 2-1 for
dimensions), allow 7.6-cm (3-in.) clearances at the sides and back
to ensure sufficient air circulation. The ultracentrifuge must have
adequate air ventilation to ensure compliance to local requirements
for vapors produced during operation.
• Relative humidity should not exceed 95% (noncondensing).
cm
in.
27.3
10.8
7.6
3.0

Figure 2-1. Dimensions of the Optima TLX Ultracentrifuge

100-V centrifuge . . . . . . . . . . . . . . . . . 90–110 VAC, 12 A, 50/60 Hz
120-V centrifuge . . . . . . . . . . . . . . . . 100–130 VAC, 12 A, 50/60 Hz
220/240-V centrifuge. . . . . . . . . . . . . . . . 198–254 VAC, 6 A, 50 Hz
73.7
29.0
7.6
3.0
53.3
21.0
58.4
23.0