Circuit Breaker Location; Thermal Breaker Location - Teledyne 10 Installation And Operation Manual

Teledyne isco, inc. installation and operation guide switch
Table of Contents

Advertisement

Circuit Breaker
Figure 4-7 Circuit breaker location
Thermal Breaker
Figure 4-8 Thermal breaker location
Risk of damage to eyesight and skin. Wear ultraviolet (UV)
light eye protection and protect skin from exposure if the
deuterium lamp is operating.
Hot surfaces may cause burns. The UV lamp and nearby sur-
faces are hot.
To reset the circuit breaker, separate the pump and column
modules (see section 2). Locate the circuit breaker illustrated in
Figure 4-7 and press the button.
The thermal breaker is located behind the waste port panel
(Figure 4-6). To reset the breaker, open the panel and locate the
reset switch (Figure 4-8). Insert a small flat blade screwdriver
into the opening and push the switch in to the right.