BioOptix™ 10
Appendix A Method Templates
A.2.2 [Blocks]
A.2.3 [Blockn]
A.2.4 [Other]
A-4
Additional rules should be observed and are described in the fol-
lowing sections.
The count in this section should equal the total number of blocks
in the template. When generating a method, the number of tabs
created will equal the count value. The acceptable value range is
1 to 10.
This section specifies what information appears on a tab. A sep-

[Blockn]

arate
must be used for each tab in the method file, incre-
n
menting
by one (Block1, Block2, Block3, ... Block10). The
value determines the tab order.
Blockn
The contents of a
• Name – This text entry is used to label the tab. Up to 20
characters can be used to provide a descriptive title.
• Type – This text entry is limited to three options which
determine the controls that will appear on the tab.
· Fluid – This type provides controls for pumping an
isocratic mix of buffer solution. An example of a Fluid
tab can be seen in Figure 3-2.
· Load – This type provides controls to pump liquid
sample onto the column, or to allow a manual
injection. An example of a Load tab can be seen in
Figure 3-3.
· Gradient – This type is used to reference a Gradient
file and provides controls for the flow rate. An
example of a Gradient tab can be seen in Figure 3-5.
• AskPause – This text entry enables the option to Pause
at End of the block. If this block in the method should
allow a choice to pause the operation, enter Yes. Enter
No to disable the choice. When disabled, the BioOptix 10
automatically proceeds to the next block.
This section contains the UV detection wavelength setting. This
value assigns a default wavelength for the method. Note that all
blocks use the same wavelength setting.
include:
n