Section 2 Installation

2.2.5 Fluid Connections

2-10
5. Connect the AC power cord to the back of the Foxy 200.
Connect the other end of the power cord to an outlet meet-
ing the electrical requirements listed on the serial number
label.
6. Connect the "Y" Serial Cable end labeled Second Instru-
ment Foxy to the RS-232-C port on the back of the Foxy
200.
7. Place the Foxy 200 to the right of the BioOptix 10 (Figure
2-1). Ensure that the full motion of the fraction collector
arm is not obstructed, especially near column positions 9
and 10 on the BioOptix 10.
Fluid line connections include Buffers, Samples, Waste, Pump
Purge, and Foxy. Tubing for the fluid connections is located in the
accessory kit, P/N 60-5209-002. The kit includes enough tubing
to complete a standard installation. Standard tubing lengths are
noted in the following instructions for each connection. This kit
also includes a tubing cutter.
Except as noted on the Foxy connections, use the following steps
for tubing connections:
1. Loosen the nut on the compression fitting.
2. Insert the tubing fully into the fitting.
3. Finger-tighten the nut.
4. Using a wrench, tighten the fitting an additional full turn.
Buffer A – These inlet ports are located on the left side of the
BioOptix 10. The ferrules on these inlet ports accept the
0.125" OD tubing, P/N 023-0503-02, found in the accessories kit.
Note
Larger or smaller outside diameter tubing will require different
ferrules.
Buffer A solutions may be unique to each channel, or a single
solution may be routed to all ten channels.
• If unique buffers will be used on each channel, cut the
tubing to length (standard length is 60" or 152 cm) and
route the tubing to the buffer solution containers.
• If a common buffer solution will be used for all ten
channels, install the Buffer Assembly (P/N
60-5204-064), found the accessory kit. Figure 2-14 illus-
trates the Buffer Assembly parts. Figure 2-15 shows the
buffer assembly installed on the BioOptix 10.