Controls and Indicators
accessories of the corresponding circuit should be checked and replaced if
necessary.
2.1.9 Output Indicator <9>
2.1.10 Patient Lead Connector <10>
Patient Lead - Mode(s) of Attachment
The colours of the connectors facilitate easy and
correct attachment of the electrodes to their
polarity.
Colour of the connector:
red = anode (+), blue = cathode (-)
Type BF component, not connected to protective ground wire.
Caution
Take care that the electrodes are by no means touched after the
current is turned up!
6
This indicator tells you to be cautious when handling the
electrodes:
Warning!
The patient lead connector <10> and the electrodes are
under voltage! Attach or remove the electrodes only
after the intensity control <6> has been turned down to
0 and the output indicator <9> is no longer flashing!
The patient lead connector <10> serves to plug in the
patient lead. You can plug plate electrodes, adhesive
electrodes or other types of electrodes into the patient
lead.
Anode (+)
vocaSTIM
Cathode (-)
-Master
®