Physiomed vocaSTIM-Master Operating Instructions Manual page 20

Table of Contents

Advertisement

Controls and Indicators
(FM)
Frequency-Modulated Current
Fixed parameters
Pulse shape: triangular
T = pulse time: 1 ms
R = release time: 70 -142 ms
Stimulation frequency: 7 -14 Hz
monophase
with galvanic basis 5%
T
R
Application:
(FaS)
Faradic Surge Current
Fixed parameters
Pulse shape: triangular
monophase
with galvanic basis 5%
T
R
Application:
14
Adjustable
Mono-/biphase
biphase
T R
Shiver-inducing frequency with automatically altered stimulation
cycles for circulatory stimulation, relaxing muscle tensions,
analgesia with little sensitive strain
Adjustable
t = pulse time: 0.1 - 10 ms (classic: 1 ms)
Mono-/biphase
T = contraction: 1 - 60 s; manual
triggering (manual release key)
R = pause: 1 – 60 s
Ramp (surge time): 4 modes
biphase
T R
Treatment of muscle atrophies, intentional exercises according to
Förster, electrogymnastics, muscle training in sports, co-ordination
exercises
vocaSTIM