Physiomed vocaSTIM-Master Operating Instructions Manual page 19

Table of Contents

Advertisement

(IG 30) P
ulse Galvanisation 30
acc. to Jantsch
Fixed parameters
Pulse shape: triangular
T = pulse time: 30 ms
R = release time: 50 ms
Stimulation frequency: approx. 12
Hz
monophase
with galvanic basis 5%
T
R
T
Application:
Circulatory stimulation, analgesia
(IG 50) P
ulse Galvanisation 50
acc. to Jantsch
Fixed parameters
Pulse shape: triangular
T = pulse time: 1 ms
R = release time: 20 ms
Surge time: 50 ms
Surge pause: 70 ms
Stimulation frequency: approx. 8 Hz
monophase
with galvanic basis 5%
T
R
Application:
Shiver-inducing frequency for circulatory stimulation, analgesia,
relaxing muscle tension, haematoma reduction.
vocaSTIM
-
Master
®
Adjustable
Mono-/biphase
biphase
T
R
T
Adjustable
Mono-/biphase
biphase
T R
Controls and Indicators
13