Therapy
automatically and will keep the set intensity as long as the manual release key
is pressed. After releasing the button, the intensity will automatically be
reduced in intervals. Pressing the button for
the time T previously selected.
Note
This safety interruption also appears in case the intensity is turned up
by more than 5.0 mA between pulses while the manual release key is
kept depressed. Whenever this message appears, slowly turn down
the intensity. The message will disappear as soon as the previously set
value is reached again.
(13) Signs of fatigue should stop the treatment (they may already occur after a few
muscle contractions)!
(14) After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
(15) Remove the electrodes.
4.6

Therapy using Medium-Frequency Current Modes

4.6.1 Pain Therapy, Hyperaemisation, Detonisation using AMF Current

Treatment Procedure
(1) Prior to attaching the electrodes, make sure that the intensity control<6> is
turned down to 0!
(2) Attach the electrodes to the respective stimulation points.
(3) Select the current mode
(4) Depending on the indication, select a frequency band or a fixed frequency.
(5) If necessary, alter the basic frequency.
(6) Set the therapy time.
(7) Set the time for treatment before turning up the intensity. Slowly turn up the
intensity (intensity control <6>). Therapy time elapses.
(8) After treatment, an acoustic signal will be issued.
Turn the intensity down to 0.
(9) Remove the electrodes.
26
If the intensity is turned up during release time, the
following message appears in the lower status bar <4>:
from the display <3>.
AMF
will trigger a current pulse of
T/R
vocaSTIM
-Master
®