Exposure Time Verification

The diaphragm 1 for panorama exposure is still adjusted at the diaphragm wheel.
1. Connect test point T1 of terminal on PCB DX1 to L of 110V power receptacle.
2. Connect common lead of pulse counter to N of 110V power receptacle.
3. Connect second measuring lead of pulse counter to test point T0.
4. Switch unit ON. Wait 1 min. for self-adjustment of the unit.
The X-ray head must be in the initial position (if not press return key R on Multitimer).
5. Remove and reinsert the film cassette.The LED above the R key on the Mutitimer must go out.
6. At the Multitimer select 72kV/10mA with P1 program.
Measurement:
7. CAUTION RADIATION ! - Depress the exposure key and hold depressed until the exposure terminates
automatically.
The pulse counter shall indicate
- - Record average pulse count.
8. Without Tele-Exposure: If specified values are obtained, switch unit OFF.
Remove meter leads and refit small metal cover (2 screws).
If specified value cannot be obtained, see Service Manual, chapter "Exposure Time Verification".
59 01 603 D3285
D3285.103.04.02.02
5.
No 1
6./ 7.
Multitimer
T1
1.
for Panorama Exposure
at 60Hz: nominal 11.3 sec.= 678 pulses±36
at 50Hz: nominal 11.3 sec.= 565 pulses±30
Maintenance Instructions
Pulse counter
3.
T0
2.
To
To
of
power receptacle
com.
N
L
110VAC
11