5.3

Error Messages

Error messages are saved in error queue which access by FIFO method. The return
first error message is the first being saved.
When the error message is over 30, the last position would be saved as -350, "Queue
overflow". The error queue can't save error message any more until there is error
message out.
When there is no error occurred, the first position would be saved +0,"No error".
-102
Syntax error
Syntax error, usually includes not allowed character symbol in command.
-108
Parameter not allowed
The device receives parameter is not allowed.
-109
Missing parameter
Parameter is missed.
-112
Program mnemonic too long
Simple command program header is over 12 characters.
-113
Undefined header
The device is received undefined header.
-114
Header suffix out of range
The header suffix is out of range.
-151
Invalid string data
Invalid string data is usually missing double quotation.
-158
String data not allowed
The device is received disallowed string data.
-170
Expression error
The device is received uncompleted parameter data, such as missing the right
parenthesis.
-222
Data out of range
The data is out of range.
-291
Out of memory
The data is out of memory.
-361
Parity error
The parity is error.
-365
Time out error
The device isn't received end character within a certain time.
-363
Input buffer overrun
The device is received over 1024 characters.
-400
Queue error
The output queue data is over 256 characters.
-410
Query INTERRUPTED
When received a query command, you don't read out the query result and then
received another query command immediately.
-420
Query UNTERMINATED
There is no data in queue, meanwhile received the command of reading output queue
data.
Remote Interface Description
5-13