Dalsa Piranha ES-80-08k40-00-R User Manual page 64

Enhanced sensitivity cameras
Table of Contents

Advertisement

64
PRNU Correction
Performing PRNU Correction
Purpose:
Syntax
N otes:
Performing PRNU to a user entered value
Purpose:
Syntax:
Syntax Elem ents:
03-032-20070-02
Perform s PRN U correction to a cam era calibrated peak value and
elim inates the d ifference in responsivity betw een the m ost and least
sensitive pixel creating a uniform response to light. For a com plete
d escription on how to use this com m and , see the Flat Field
Correction Overview on page 58.
ccp
Perform all analog ad ju stm ents before calcu lating PRN U.
Perform FPN correction before PRN U correction.
If FPN cannot be calibrated , u se the
coefficients to zero, and save them to m em ory w ith the
com m and . You can then ad ju st the d igital offset (
com m and ) to rem ove som e of the FPN .
ccp
The
com m and is not available w hen the cam era shift
d irection is externally controlled . Direction control m u st be
stable w hile the cam era is calcu lating coefficients (see
Direction Control on page 26).
Ensu re cam era is operating at its expected analog gain,
integration tim e, and tem peratu re.
To avoid losing you r cu rrent d irection coefficients, you m u st
save the PRN U coefficients u sing the com m and
changing cam era shift d irection or changing from TDI to
Area Mod e.
Available in TDI Mod e only.
Execu ting these algorithm s cau ses the ssb com m and to be
set to 0 (no backgrou nd su btraction) and the ssg com m and
to 4096 (u nity d igital gain). The pixel coefficients are d isabled
(epc 0 0) d u ring the algorithm execu tion bu t retu rned to the
state they w ere prior to com m and execu tion.
Perform s PRN U calibration to user entered value and elim inates the
d ifference in responsivity betw een th e m ost and least sensitive pixel
creating a uniform response to light. Using this com m and , you m ust
provid e a calibration target.
Executing these algorithm s causes the ssb com m and to be set to 0
(no background subtraction) and the ssg com m and to 4096 (unity
d igital gain). The pixel coefficients are d isabled (epc 0 0) d uring the
algorithm execution but returned to the state they w ere prior to
com m and execution.
cpa i i
i
PRN U calibration algorithm to use:
1 = This algorithm first ad justs each tap's analog gain so that 8-
13% of pixels w ithin a tap are above the value specified in the
target value param eter. PRN U calibration then occurs using the
peak pixel in the region of interest.
This algorithm is recom m end ed for use only w hen FPN is
negligible and FPN coefficients are set to zero. Since this
Piranha ES-xx User Manual
rpc
com m and to reset all
wpc
sdo
wpc
before
Teledyne DALSA

Hide quick links: