The steriliser is not detecting
the SD memory card, or the
SD card is write-protected.
W85
The conductivity read by the
conductivity meter on the
autoclave is at the limits of
the acceptable values.
The conductivity read by the
conductivity meter on the
autoclave is at the limits of
the acceptable values.
Door
W43
faulty.
Door lock electromagnet out
W44
with door open.
Lower
W80
temperature not suitable for
cycle start.
Upper band heater
W81
temperature not suitable for
Upper band heater
W82
temperature not suitable for
W90
temperature sensor faulty.
Upper band temperature
W91
Lower band temperature
W92
Faulty condensation battery
W93
W94
Pressure sensor faulty.
Chamber temperature
W95
lock
electromagnet
band
heater
cycle start.
NGV cycle start.
Steam generator
sensor faulty.
sensor faulty.
sensor.
sensor faulty.
81
Check that the SD memory card is present
and
correctly
installed.
steriliser, remove the SD memory card and
check that the safety switch allows writing.
Empty the clean water tank as soon as
possible, and refill it with demineralised or
distilled water of the best quality.
Empty the clean water tank, and refill it with
demineralised or distilled water of the best
quality.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Manually move the electromagnet back in by
pushing the pin toward the steriliser.
Resistance fault.
Resistance fault.
Resistance fault.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.