Euronda E9 Next 18L Instruction Manual page 80

Table of Contents

Advertisement

Faulty condensation
E44
battery temperature
sensor.
Chamber temperature
E45
sensor faulty.
Pressure sensor
E46
faulty.
Door closed sensor
E47
faulty.
Door lock sensor
E48
faulty.
Steam generator
E51
inactive.
Temperature of
E54
steam generator too
high.
Temperature of upper
E55
band too high.
Temperature of lower
E56
band too high.
Temperature of lower
band heater too high
E58
in the pressure
maintenance cycle.
Condensation battery
E59
temperature too high
Problems about write
E60
on Sd card
Water
injections
E62
finished.
Water not being
E81
supplied by the
Aquafilter® deioniser.
Comunication
problem between
E99
display board and
power board
Fault in the sensitive
element or in the
probe connection.
Fault in the sensitive
element or in the
probe connection.
Fault in the sensitive
element or in the
probe connection.
Fault in the door
closed position switch.
Fault in the door
locked position switch.
Fault in the steam
generator or electronic
board or steam
generator safety
thermostat tripped.
Fault in the electronic
board or temperature
sensor.
Fault in the electronic
board or temperature
sensor.
Fault in the electronic
board or temperature
sensor.
Fault in the electronic
board or temperature
sensor.
Sensor, connection, or
fan fault.
SD card inserted after
power on or removed
during a cycle
Boiler overloaded or
clogged with scale or
water injection pump
inefficient.
Error in the hydraulic
or electrical
connections with
Aquafilter or electronic
fault.
79
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Switch the steriliser off and then back on
again. If the problem persists contact the
assistance service.
Open and close the door a few times. If the
problem persists contact the assistance
service.
Perform a sterilisation cycle. If the problem
persists contact the assistance service.
Contact the assistance service.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Switch off the steriliser and allow it to cool
for a few hours, then perform a sterilisation
system. If the problem persists contact the
assistance service.
Contact the assistance service.
Switch off the autoclave and check if the
sd card is in the right place. If the problem
persists contact the assistance service
Perform a sterilisation cycle with a small
load in the boiler (just one tray). If the
problem persists contact the assistance
service.
Make sure that the Aquafilter connections
are correct and that there are no crushed
or bent pipes. Make sure the Aquafilter
inlet tap is open. If the problem persists
contact the assistance service.
Switch off the steriliser and after a few
seconds turn it on again. If the problem
persist contact the assistance service
E9_Next_Ing_rev03 – 10/10/2018
Aquafilter
E9 NEXT
®