Description Of Tests; Appendix 6 - Euronda E9 Next 18L Instruction Manual

Table of Contents

Advertisement

Light & Stock Program (S121 and S134)
With this program it is possible to sterilize only solid single wrapped instruments (not hollow).
You can sterilize at maximum 4,5 kg. with the 18 lt. model, and 6 kg. with the 24 lt. version.

APPENDIX 6

It is important to periodically verify the performance of the unit by performing the appropriate tests;
E9 can perform three different ones:
-
B&D test
-
Vacuum test
-
Helix test
The parameters of the respective cycles are as follows:
Parameter
Temperature
Pressure
Length of sterilization phase
(plateau period)
Drying time
Total time
Vacuum test
The sole purpose of the diagram is to illustrate the quality performance of the cycle
This test is performed in order to check the performance of the unit, in particular:
the efficiency of the vacuum pump;
-
the seal of the hydraulic circuit.
-
The cycle is structured as follows:
1. a vacuum is created to the minimum pressure value indicated in the load pre-treatment phase
2. this pressure is maintained for 5 minutes and then measured
3. this pressure is maintained for 11 minutes and then measured
In compliance with EN13060, the test requires a tightness test of less than or equal to 1.3 mbar/min during
the 10 minutes of test; if the leakage is greater than this value, the outcome of the test is negative; the seal
of the hydraulic circuit of the device must be checked.

Description of Tests

Cycles
VACUUM
NEXT 18/24L
--------
minimum pressure
--------
--------
B&D
NEXT 18/24L
135.5°C
2.15 bar
3'30''
--------
21'
26'
32'
30'
71
Aquafilter
E9 NEXT
HELIX
NEXT 18/24L
135.5°C
2.15 bar
3'30''
--------
26'
33'
E9_Next_Ing_rev03 – 10/10/2018