A sterilisation or test cycle can be manually stopped at any time.
DO NOT INTERRUPT THE CYCLE BY SWITCHING OFF THE POWER SUPPLY TO THE UNIT
as this may cause damage. Always use the manual stop procedure indicated in this paragraph.
To perform a manual stop, press the RED HAND button in the Cycle in Progress screen, and then the RED
HAND button in the next screen. This procedure applies to all sterilisation and test cycles.
At this point, the steriliser will start a sequence of operations allowing the steam to be safely discharged and
return the boiler pressure to the external level.
7.4.1 Manually stopping a cycle before or during the sterilization phase
If a cycle is stopped before the sterilisation phase has terminated, the load in the boiler must be considered
as NOT STERILE. After the manual stopping operations, an error screen appears on the LCD display. The
door is locked. Touch the screen to release it.
7.4
MANUALLY STOPPING A CYCLE OR A TEST
37
Aquafilter
E9 NEXT
E9_Next_Ing_rev03 – 10/10/2018