Before performing any maintenance operations, carefully read the following safety
instructions and, especially, chap. 3 "Safety".
CAUTION:
to only use ORIGINAL SPARE PARTS.
HAZARD: HIGH INTERNAL VOLTAGES.
CAUTION: DISCONNECT THE POWER SUPPLY BEFORE STARTING WORK. Non-
observance may cause serious injury to people or may seriously damage the unit.
ALL MAINTENANCE OPERATIONS MAY ONLY BE PERFORMED BY THE RESPONSIBLE AUTHORITY
OR BY THE TECHNICIANS AUTHORISED BY THE ASSISTANCE SERVICE OF EURONDA S.p.A.
-
Observe the intervals prescribed or indicated in this manual.
-
It is forbidden to remove the safety devices installed on the machine (see chap. 3.3 "Safety devices").
Check them at regular intervals.
-
If an effective danger situation arises, press the ON-OFF button (4 of Fig. 5.1.1-1) immediately.
-
Unauthorised people must stay at a safe distance from the machine during maintenance operations.
Just like all electric units, this unit must be correctly used, serviced, and checked at regular intervals. These
precautions will ensure continuous, safe, and efficient use of the unit.
To prevent operator hazards, the unit must be subject to regular checks and servicing by the technical
assistance service.
-
For good preservation of the unit, periodically clean all external parts with a soft cloth damped with normal
neutral detergent (do not use corrosive or abrasive products).
-
Do not use abrasive cloths, pads or metal brushes (or anything abrasive) to clean the metal.
-
Before starting each cycle, clean the door seals carefully using a damp cloth.
-
The formation of white stains on the base of the chamber shows that the demineralised water used is of
poor quality.
Maintenance programme
FREQUENCY
DAILY
WEEKLY
ANNUALLY
EVERY 500 cycles
EVERY 1000 cycles
EVERY 1000 cycles/2 years 1000 cycle maintenance kit replacement
EVERY 1500 cycles/3 years Replacing the bacteria filter and the seals.
EVERY 2000 cycles/4 years 2000 cycle maintenance kit replacement

CHAPTER 9

9.1

SAFETY WARNINGS

when replacing components that directly or indirectly affect safety, it is essential
9.2

ORDINARY MAINTENANCE

OPERATION
Cleaning of the door seal.
General cleaning of the external surfaces.
General cleaning of the internal surfaces.
Cleaning of the sterilization chamber.
Cleaning of the trays and the support.
Maintenance of the safety valve.
Replacement of the bacteriological filter
Replacement of the seals
56
Aquafilter
E9 NEXT
E9_Next_Ing_rev03 – 10/10/2018