The following table shows the warning messages given by the steriliser using symbols or codes when it
detects a problem that prevents a cycle from starting.
CODE
W41
W84
DESCRIPTION
Used water at maximum
level.
Clean water under minimum
level.
An attempt was made to
start a cycle with the door
open.
Steriliser too high.
The conductivity read by the
Aquafilter
®
deionizer is out of
range and the automatic
supply of water is therefore
impossible.
80
Switch off the steriliser and after a few
seconds turn it on again. If the problem
persist contact the assistance service
SOLUTION
Empty the used water tank.
Fill the clean water tank with distilled or
deionized water.
Before running a cycle, close the door.
The steriliser temperature is too high for the
requested operation. Switch it off and leave it
to cool leaving the door open.
Replace the cartridges in the Aquafilter
deionizer.