EVERY 2500 cycles/5 years Replacing the bacteria filter and the seals.
EVERY 3000 cycles/6 years 3000 cycle maintenance kit replacement
EVERY 3500 cycles/7 years Replacing the bacteria filter and the seals.
EVERY 4000 cycles/8 years 4000 cycle maintenance kit replacement
AFTER 10 YEARS
WHEN NECESSARY
Cleaning the sterilization chamber, accessories, door and seal
Sterilization chamber
Clean the sterilisation chamber thoroughly (Fig. 9.2-1), after having removed the tray support, using a non-
abrasive damp cloth.
To dampen the cloth, use only and exclusively distilled or demineralised water. Follow the same procedure
for cleaning the trays and their support. Cleaning the sterilization chamber is important for eliminating
deposits that could compromise the good working order of the machine. To remove the tray support: remove
the support from the chamber (Fig. 9.2-2) and, after finishing cleaning, refit it following the same procedure in
the reverse order.
CAUTION: DISCONNECT THE POWER SUPPLY BEFORE STARTING WORK. Non-
observance may cause serious injury to people or may seriously damage the unit.
CAUTION: take care not to damage the probe at the bottom of the chamber.
Fig. 9.2-1
Do NOT use disinfecting substances to clean the chamber.
Request a structural check of the chamber.
Adjustment of the closing mechanism.
57
Fig. 9.2-2
E9_Next_Ing_rev03 – 10/10/2018