APPENDIX 9

Euronda E9 NEXT is equipped with a system for controlling all the components of the unit; when the Process
Controller detects a fault either in a component or in the overall behaviour of the machine, an error message
appears, preceded by an intermediate "Please wait" screen, necessary for the machine to perform its pressure
levelling.

Troubleshooting

END OF CYCLE AND ERROR SCREEN
This screen appears after a manual stop before the end of the sterilisation cycle,
or when an operation error has caused the unit to stop the cycle, and therefore
the load is not sterile. The screen shows the name of the cycle at the top and the
relative error code at the bottom (in this case) and the door locked icon (padlock).
If this screen error appears due to a manual stop by the user, the error code is
not used (like in this case).
The door is locked. To release it, touch the central part of the screen. The handle
chamber shines red.
LOAD WET END-OF-CYCLE SCREEN
This screen appears when a cycle is interrupted by a manual stop ordered by the
user, after the end of the sterilisation phase: the load in the boiler is sterile, but
the drying process has not been completed. The load can therefore only be used
immediately and cannot be preserved. The door is locked. To release it, touch
the central part of the screen.