Table 6. Recommended Separation Distances - Nonin WristOx2 3150 Operator's Manual

Wrist-worn pulse oximeter with bluetooth wireless technology
Hide thumbs Also See for WristOx2 3150:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
This table details the recommended separation distances between portable and mobile RF
communications equipment and this device
This device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. Users of this device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communication equipment (transmitters) and the device as
recommended below, according to maximum output power of the communications equipment.
Rated Maximum Output
Power of Transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTES:
1. At 80 MHz and 800 MHz, the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
40
Table 6: Recommended Separation Distances
Separation Distance According to Frequency of Transmitter
150 kHz to 80 MHz
d
=
1.17 P
0.12
0.37
1.2
3.7
12
80 MHz to 800 MHz
d
=
1.17 P
0.12
0.37
1.2
3.7
12
Technical Information
800 MHz to 2.7 GHz
d
=
2.33 P
0.23
0.73
2.3
7.3
23