Memory and Data

Memory and Data
The WristOx
Model 3150 measures, collects, and stores up to 1,080 hours of SpO
and pulse
2,
2
rate data with a 4-second data collection rate. Data collected at a 1 or 2-second rate reduces
memory capacity to 270 or 540 hours, respectively.
When the memory is full, the device overwrites the oldest existing data with the new data. Each
time the device is turned on, data are automatically stored in memory. Data collection of less
than 1 minute is not retained in memory.
Each time the device turns on, the current oximeter time and date (if the clock is set properly)
are stored in memory to allow quick differentiation of recording sessions. Patient SpO
and
2
pulse rate are stored every 4 seconds (default), or every 1 or 2 seconds if programmed using
nVISION software. The oxygen saturation values are stored in 1% increments in the range of
0 to 100%.
This device contains non-volatile memory. Removing or replacing batteries does not affect the
data stored in memory. Stored data remains in memory until overwritten by newer data or
cleared from memory with nVISION software.
NOTE: Downloading data in memory does not clear memory. To clear memory, see
"nVISION Settings."