Playing Back Memory Data - Nonin Avant 9600 Operator's Manual

Digital pulse oximeter
Hide thumbs Also See for Avant 9600:
Table of Contents

Advertisement

C o m m u n i c a t i o n

Playing Back Memory Data

The Avant 9600 has a Memory Playback feature, allowing stored data to be output
through the RS-232 serial connection.
1.
With the unit off, connect the RS-232 connector port of the Avant 9600 to the
back of your computer using a null modem cable.
2.
With the Avant 9600 still off, press and hold the Time/Date button while
pressing the ON/STANDBY button.
• All LEDs will illuminate briefly. PLy bac will appear in the
SpO
the Avant 9600 is in Playback mode.
3.
The PLy bAC message will disappear when memory playback is complete.
(Memory playback may take up to 8 minutes, depending upon the amount of
data.) Pressing the ON/STANDBY button will exit Playback mode.
4.
A CLr no message will be displayed, and three informational tones will
sound.
5.
(OPTIONAL): To clear the memory:
• Use the Plus or Minus buttons to select Clr YES.
• Press the ON/STANDBY button.
• To confirm the clearing of memory, use the Plus or Minus buttons
to select DEL YES.
• Press the ON/STANDBY button again.
• dnE CLr confirms that the memory is clear.
6.
Press the ON/STANDBY button to return to normal operation.
NOTE: Patient memory cannot be cleared when the Avant 9600 is in
Patient Security mode.