Model CAT100
c.
Power Supply
The CAT 100 power supply is located in
the base of the explosion proof enclosure,
beneath the analyzer. Removal of the
analyzer (Section 5-2a, page 5-3) is nec-
essary to service the power supply.
Neutral
POWER
Hot
Rosemount Analytical Inc.
AC
Figure 5-26. Power Supply Connections
A Division of Emerson Process Management
Test Point:
Test Device:
Signal:
Failure:
Possible rea-
sons:
TP2
TP1
Instruction Manual
748441-D
April 2002
TP1 (+ )
TP2 (gnd)
DVM
24 V DC (± 1 V)
No voltage or incor-
rect voltage
Faulty power supply
(replace)
EMI Board
Fuses
Maintenance and Service
5-37