3020M5001_2

7.6.4 Editing a Configured Channel

Note: There was two kinds of timeout in menu mode. One is menu timeout and the other is authentication
time out. If there is no touch input, then it will move to upper menu and then finally back to Channel Status
screen. The menu timeout can be configured in Configuration > General > Timeout. The default menu timeout
is 90 seconds. The authentication timeout is to force the user to log out automatically when there is no activity
for the configured duration. If the user logged out automatically due to authentication timeout, a user will be
asked to log in again.
To edit an already configured channel:
1) Login as an Service or higher access level
2) Select the Menu > Configuration > Channel Settings and then select an IO module.
3) Select the required channel.
4) Select the screen that you want to edit (refer to previous sections of this manual).
5) Enter the required values in the field New Value and touch 'Finish'.
6) Review all parameters are valid and touch 'Apply'.
7) The system will return the new value; touch 'Update' to confirm.
Note: Refer to Ch.14 Configurable Parameter Reference Guide for the list of available detectors. You can also
use the TPPL Sensor Catalog to select listed gases, or you can enter 'User gas' details if respective gas is
not listed.
Refer to Ch.8.6 How to Update the Sensor Catalogue for further information.
Commissioning
When selecting a new target gas for units the sensor must be recalibrated.
63
WARNING
Detector Name
Sensor Type
Touchpoint Plus Wireless
Technical Handbook