Transducer Cable Removal And Installation - Emerson Daniel 3410 Series Instructions For Use, Maintenance And Installation Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Maintenance and Troubleshooting manual
P/N 3-9000-769
12. Hold transducer capsule vertically and apply a small amount of Acoustic Coupling
13. Carefully spread the coupling fluid on the surface using the tip of the plastic bottle
14. Wait for 30 - 60 seconds to let the coupling fluid spread more evenly on the surface.
15. Insert the transducer capsule into the transducer stalk while aligning the anti-
16. Make sure that the indicator ring on the transducer capsule is flush with the end of
17. Hand-tighten the transducer retainer onto the stalk.
18. Follow
3.8

Transducer cable removal and installation

Daniel 3410 Series Ultrasonic Gas Flow Meters have blue transducer cables and a
transducer cable nut that threads directly into the back of the transducer holder. The
cables use plastic glands that are installed on the meter's Base Enclosure.
Note
Make a note of the exiting cabling path layout to allow proper tie-wrap configuration later
in this procedure.
Maintenance and Troubleshooting manual
c) Turn the bottom piece counter-clock wise by 90 ° to let the tap snap into the
correct slot. Make sure the tap is secured in the slot.
Fluid (P/N: 1-360-01-650) to the surface of Kapton tape.
for the coupling fluid.
Note
Try to avoid pressing the Kapton tape while spreading the coupling fluid.
rotation pin on the capsule with the slot on the inner surface of the stalk.
the transducer stalk to ensure the capsule is set to the correct length (see
3-3).
Section 3.5.2
to install the transducer assembly in a meter body.