Troubleshooting
June 2019
This manual may be downloaded from the following location:
documents/automation/user-s-manual-en-104978.pdf
For troubleshooting communications with AMS Device Manager, refer to the help
documentation and support at the following web site:
catalog/asset-reliability/ams-ams-device-manager
2.2
Troubleshooting the meter
Table 2-1
firmware or connections and recommended actions to resolve the problem(s).
Table 2-1: Troubleshooting
Error
Acquisition Module Error
Acquisition Module is not compatible with
firmware
24
and the following sections show errors that may occur with the meter hardware,
Maintenance and Troubleshooting manual
https://www.emerson.com/
https://www.emerson.com/en-us/
Recommended action(s)
•
Check interconnect cable between
Acquisition Module and the CPU Module.
•
Attempt the Program Download procedure
to install the firmware.
—
Cycle power to the meter.
—
Replace the Acquisition Module.
—
If the Acquisition Module cannot be
reprogrammed, collect a complete
Archive log and contact your local area
Daniel service representative.
•
Upgrade the firmware in the meter to the
latest version using Daniel MeterLink.
Contact your Daniel service representative
to obtain the latest firmware.
•
Replace the Acquisition Module.
Gas Ultrasonic Flow Meters
P/N 3-9000-769