Maintenance and Troubleshooting manual
P/N 3-9000-769
Procedure
1. Blow the line down according to the site standard operating procedures.
2. Ensure that the line pressure is down to atmospheric pressure prior to disassembly.
3. Disconnect transducer cabling and chordset from the transducer assembly by
4. Disconnect the transducer cable nut. Pull the chordset from the transducer
5. If the meter is equipped with T-21 or T-22 transducers, remove the transducer
Maintenance and Troubleshooting manual
turning the cable nut counter-clock wise.
assembly.
Note
Do not twist the chordset as you disconnect it from the transducer assembly.
retainer using a 1 1/8" wrench and then disconnect and remove the transformer
(Figure
3-5).
Note
T-21 transducers use W-01 transformers and T-22 transducers use W-02
transformers.
Meter repairs
June 2019
59