Replace The Meter Electronics - Emerson Daniel 3410 Series Instructions For Use, Maintenance And Installation Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Maintenance and Troubleshooting manual
P/N 3-9000-769
24. Twist and adjust wire removing all slack and seal. Remove excess wire (see
25. If required by the site operations manager, have an electrician fully test the
26. Power down the system and apply the sealing compound to the conduit and allow
3.9

Replace the meter electronics

The following procedure should be performed by a qualified service technician or trained
personnel. Observe all warning labels on the meter before starting this procedure.
The Daniel 3410 Series Gas Ultrasonic Flow Meter Transmitter Electronics Enclosure
consists of the following:
CPU Module assembly (P/N 1-360-03-010)
Optional I/O Module (RS-232 or RS-485)
I.S. Barrier Board (P/N 360-03-004)
Power Supply (P/N 360-03-003)
Backplane Board (P/N 360-03-007)
The Daniel 3410 Series Gas Ultrasonic Flow Meter Base Enclosure consists of the following:
Acquisition Module (P/N 1-360-03-008)
Acquisition Cable (P/N 360-01-595)
Should the Daniel 3410 Series Ultrasonic Gas Flow Meter require disassembly in the field
(i.e., check boards, change switch settings, or replace boards), to prevent electrostatic
damage to the electronic boards, always use a ground strap while handling the circuit
boards. If one is not available, make sure you are electrically discharged before touching
the boards by first touching a metal surface such as a ground lug on the meter or a table.
Maintenance and Troubleshooting manual
3-23).
connections. After the Acceptance Test is witnessed and approved, seal the
conduit.
to set in accordance with manufacturer specifications.
Meter repairs
June 2019
Figure
87

Advertisement

Table of Contents
loading

This manual is also suitable for:

341434123411Daniel 3414Daniel 3412Daniel 3411

Table of Contents